"***Vipat REF H0020-100***Lot 100533***2015-04***STERILE EO Elastomeric pump***2mL/Hr***100mL***Medpro International (Thailand) Ltd. 360, Moo 6, Chonburi Industrial Estate, Tambon Bowin, Amphur Sriracha, Chonburi 20230, Thailand***Victus Inc. 4918 SW 74 Court 33155 44400***" Packed 24 units per case. The devices are intended for patients requiring intravenous, percutaneous, subcutaneous, intra-operative sites or epidural administration of medications.

This recall has been terminated (originally issued October 24, 2011).

Company
Victus, Inc.
Recall Initiated
September 20, 2011
Posted
October 24, 2011
Terminated
January 12, 2012
Recall Number
Z-0080-2012
Quantity
63 units
Firm Location
Miami, FL
Official Source
View on FDA website ↗

Reason for Recall

Victus, Inc. Miami, FL is recalling Vipat Brand Elastomeric Infusion Pump, Model H0020-100, Lot 100533. The product may have a higher flow rate than specified. The use of this product may lead to over-administration of drug solutions to patients.

Distribution

USA (nationwide) including Puerto Rico and the states of FL and PA.

Lot / Code Info

Lot 100533

Root Cause

Nonconforming Material/Component

Action Taken

Victus, Inc. sent an "URGENT: MEDICAL DEVICE RECALL" letter dated July 15, 2011 to affected customers. The letter describes the product, problem, and actions to be taken by the customers. The letter instructs customers to discontinue use, remove the products from inventory and quarantine the products. Additionally, the firm indicated that their customer service department would contact customers to coordinate the return of affected products. Customers are directed to call 305-663-2129 ext. 110 for questions regarding this notice.