Welch Allyn AED 10 Defibrillator; a battery powered automated external defibrillator; Manufacturer: MRL, Inc., A Welch Allyn Company, 100 Asbury Drive, Buffalo Grove, Illinois 60089; Model AED10, part numbers 970300 (defibrillator) and 970302 (defibrillator with soft shell carrying case)

Company
Mrl, Inc., A Welch Allyn Company
Recall Initiated
April 26, 2004
Posted
July 20, 2004
Terminated
November 15, 2004
Recall Number
Z-0918-04
Quantity
460 units
Firm Location
Buffalo Grove, IL

Reason for Recall

A faulty component has been identified which could result in an intermittent shock button (button #3) failure on the AED10 defibrillator.

Distribution

Nationwide, and internationally to the United Kingdom, Ireland, Singapore, France and South Africa.

Lot / Code Info

part numbers 970300 and 970302; serial numbers 4613, 4888/4891, 4897, 6369/6386, 6396/6704, 6706, 6709/6743, 6745/6803, 6806, 6807, 6809, 6811, 6813/6832, 6836/6838, 6853/6855, 6912, 6914, 6917, 6921, 6922, 6924/6926

Root Cause

Other

Action Taken

Urgent Medical Device Recall letters dated 4/26/04 were sent to the end users via certified mail on 4/26/04. The letters informed the users of the intemittent shock (fire or 3) button failure on the AED 10. The letters included a list of the affected serial numbers, and requested the users to call Welch Allyn MRL Technical Support as soon as possible at 1-800-462-0777 for a replacement AED 10 and to answer any questions. A replacement AED 10 will be sent to the end users within 10 days, and the users were requested to return their AED 10 in the same shipping box accompanying the replacement unit. A follow-up letter will accompany the replacement unit with instructions on how to verify the affected units, packing instructions for the return of the affected units to Welch Allyn MRL, and a correction response form to be completed and enclosed with the returned unit.