RecallDepth

ProPlan CMF Guide(s) Model(s) REF SD900.008 Description: Single Use Only : Patient-Specific Guide, Orthognathic, Final Single Use Only Rx only. Used as surgical tools to transfer a pre-operative plan to surgery. .

This recall has been terminated (originally issued December 26, 2013).

Company
Materialise Usa Llc
Recall Initiated
November 8, 2013
Posted
December 26, 2013
Terminated
January 7, 2014
Recall Number
Z-0578-2014
Quantity
1
Firm Location
Plymouth, MI
Official Source
View on FDA website ↗

Reason for Recall

Patient Specific Guide contained the incorrect case report. The case report contains patient specific surgical plan information.

Distribution

Distributed in Canada.

Lot / Code Info

Item reference SD900.106; Unique Instrument ID MC13-OVA-OGO_2

Root Cause

Labeling Change Control

Action Taken

Customer/ ordering physician was made aware of the issue11/08/2013. Revised accurate case report was immediately provided.