OraQuick Ebola Rapid Antigen Test, Product Code 3001-2807; Intended for the qualitative detection of antigens for Ebola viruses.

This recall has been terminated (originally issued August 17, 2016).

Company
Orasure Technologies, Inc.
Recall Initiated
July 20, 2016
Posted
August 17, 2016
Terminated
May 10, 2017
Recall Number
Z-2403-2016
Quantity
3,075 units
Firm Location
Bethlehem, PA
Official Source
View on FDA website ↗

Reason for Recall

Failed stability testing at 8 months. This lot produced intermittent false negative results for the positive kit control with devices stored at 30C conditions.

Distribution

Distributed to one customer only in US state of Georgia.

Lot / Code Info

Product Code 3001-2807 and the lot number is 6648965. Expiration date October 31, 2016.

Root Cause

Process control

Action Taken

OraSure contacted the sole customer via email and mail (letter dated July 21, 2016) to request return of all the product from the affected lot.