Sedecal SP-HF 4.0 Portable X-Ray System; Manufactured by SEDECAL, C/Pelya 9 Pol. Ind. Rio De Janeiro, 28110 Algete (Madrid) Spain

Company
Sedecal Usa, Inc.
Recall Initiated
October 27, 2003
Posted
October 29, 2003
Terminated
March 24, 2004
Recall Number
Z-0001-04
Quantity
1 unit
Firm Location
Arlington Heights, IL

Reason for Recall

A minimum source-skin-distance of less than 30 cm and not identifying the tube manufacturer on the tube housing label resulted in the SP-HF-4.0 Portable Systems not complying with the U.S. Federal performance standard.

Distribution

Vermont

Lot / Code Info

serial #PK10058

Root Cause

Other

Action Taken

An 'Important Safety Notice' dated 10/27/03 was sent to the one direct account on the same date, informing them of the non-compliances with the performance standard and requesting them to call Sedecal at 800-920-9525 to arrange for the modification of the x-ray unit at no cost to the user.