IMx Tacrolimus II Calibrators, list 3C10-10; the pack contains six 4.5-mL bottles, with Calibrator A having 0 ng/mL, Calibrator B having 3 ng/mL, Calibrator C having 6 ng/mL, Calibrator D having 12 ng/mL, Calibrator E having 20 ng/mL and Calibrator F having 30 U/mL; Abbott Diagnostics, a Division of Abbott Laboratories, Abbott Park, IL 60064

Company
Abbott Laboratories Hpd/Add/Gprd
Recall Initiated
September 19, 2003
Posted
October 24, 2003
Terminated
October 29, 2004
Recall Number
Z-0063-04
Quantity
2677 kits
Firm Location
Abbott Park, IL

Reason for Recall

Invalid test results are being generated due to calibration errors when using calibrator lots 94165M100 and 94164M100.

Distribution

Nationwide and internationally to Guatemala, Venezuela, Peru, Argentina, Dominican Republic, Canada, Japan, Singapore, Thailand, Korea, Australia, Costa Rica, Panama and Great Britain.

Lot / Code Info

list 3C10-10, lots 94165M100 and 94164M100

Root Cause

Other

Action Taken

Recalled by letters dated 9/19/03 and 10/7/03. The accounts were informed of the inability to calibrate the IMx Tacrolimus II assay when using calibrator lots 94165M100 and 94164M100, and initially requested those accounts experiencing an increase in calibration-related errors to report the issue to Abbott for a replacement calibrator kit. The 10/7/03 letter requested the accounts to discontinue use and destroy any of the two affected calibrator lots found in inventory.