Medline ReNewal Reprocessed ViewFlex Xtra ICE Catheters: ViewFlex Xtra ICE Catheter 9F x 90cm D087031RH 20197344019554

Recall Initiated
December 22, 2025
Posted
February 6, 2026
Recall Number
Z-1323-2026
Quantity
650
Firm Location
Northfield, IL

Reason for Recall

These lots of reprocessed catheters may contain small particles of residual material. If affected devices are used, there is a risk of inflammatory response, systemic infection, or granulomatous reaction or thrombus reaction which may lead to cerebral or pulmonary embolism, or deep vein thrombosis. Expansion of recall Z-2610/2614-2025

Distribution

Distribution US nationwide.

Lot / Code Info

UDI-DI (case) 20197344019554¿ (ea) 10197344019557 Lots EP250324 EP250416 EP250512 EP250521 EP250609 EP250630 EP250710 EP250724

Root Cause

Under Investigation by firm

Action Taken

On December 22, 2025, the firm issued Urgent Medical Device Recall letters as a follow-up to an earlier action in 2025 for the same issue, but newly identified item numbers and lots of previously identified item numbers. Notably, the products associated with the new/updated action are to be destroyed by the customer on-site for credit, and not returned to the manufacturer.