Decompression Table Model E9011. This device applies cervical and/or lumbar decompression with practitioner definable target treatment location, treatment pounds, and patterns of treatment.

This recall is currently active, issued July 10, 2024. It was issued by Pivotal Health Solutions, Inc..

Company
Pivotal Health Solutions, Inc.
Recall Initiated
April 29, 2024
Posted
July 10, 2024
Recall Number
Z-2326-2024
Quantity
32 tables
Firm Location
Watertown, SD
Official Source
View on FDA website ↗

Reason for Recall

This product is being recalled due to potentially insufficient rework on a part (E7130-B) on the lumbar section of the frame on 6 distributed tables. There were 4 additional tables with the faulty E7130-B part that were manufactured, but had not been distributed yet.

Distribution

Distributed to CA, FL, TX, ID, GA, and TN

Lot / Code Info

UDI-DI 00850008449826; Model Number: E9011; Serial Numbers: WTN-DOC03132457209 WTN-DOC-3132457210, WTN-DOC03132457235, WTN-DOC03132457237, WTN-DOC03132457238, WTN-DOC03132457239 Expanded Recall serial numbers: WTN-DOC04262458492, WTN-DOC04262458490, WTN-DOC04262458487, WTN-DOC04262458488, WTN-DOC04262458489, WTN-DOC0332457239, WTN-DOC04082457899, WTN-DOC04262458484, WTN-DOC03132457238, WTN-DOC03132457237, WTN-DOC03132457210, WTN-DOC04082457897, WTN-DOC04082457898, WTN-DOC03132457201, WTN-DOC04192458307, WTN-DOC04082457896, WTN-DOC03132457209, WTN-DOC03132457235, WTN-DOC04262458496, WTN-DOC03132457240, WTN-DOC03132457241, WTN-DOC04082457882, WTN-DOC04172458185, WTN-DOC04082457893, WTN-DOC04082457894, WTN-DOC04082457895

Root Cause

Nonconforming Material/Component

Action Taken

Consignees were initially notified of the recall via phone call starting on 4/29/2024. A followup email or certified mail were sent to the consignees on 6/11/2024. Customers were instructed that Fed Ex was contacted for pickup of the table so the table can be packaged and returned to Pivotal Health Solutions at 3003 9th Ave. SW, Watertown, SD 57201. A recall expansion led to a second updated recall notification being sent to consignees beginning on June 26, 2024 via email. This letter instructs the consignees to stop using the tables listed in the letter. The firm will be contacting the consignees to schedule an engineer to come out to the consignee site to inspect the table(s). The table(s) will be replaced in found to contain an affected part. The consignee is also asked to return a response form.