RecallDepth

Synergetics I-Pack Injection Kit with Drape -For use with Ophthalmic injections Model/Catalog Number: 18061

This recall is currently active, issued July 11, 2024. It was issued by Microspecialties, Inc..

Company
Microspecialties, Inc.
Recall Initiated
June 5, 2024
Posted
July 11, 2024
Recall Number
Z-2334-2024
Quantity
96 units
Firm Location
Middletown, CT
Official Source
View on FDA website ↗

Reason for Recall

Could not confirm products were exposed to the proper sterilization cycle can potentially cause post operative wound infection or injection site infection.

Distribution

MO

Lot / Code Info

UDI: 10841305101071 LOT: P63044983R

Root Cause

Release of Material/Component prior to receiving test results

Action Taken

MICROspecialties issued Voluntary recall for Synergetics Eye Trays letter on 6/3/24 to the Distributor Synergetics. Letter states reason for recall, health risk and action to take: Review: Inventory for affected product codes Communicate with customers impacted by this notice. Segregate and Quarantine all affected product. Utilize return instructions below Disseminate this notice to any customers who may have been affected. Return the below acknowledgement form by Fax to:860-788-2030 Or email to: quality@microspecialties.com Contact quality@.microspecialties.com for product return instructions. Please return by 6/25/24. For additional information regarding this please contact quality@.microspecialties.com.

More recalls by Microspecialties, Inc.

View all recalls by this company →