Prelude Guide Sheath Roadster, REF PG5F45S018, 5F 0.076 (1.9 mm), 45 cm, STERILE EO, RX ONLY
This recall is currently active, issued May 20, 2024. It was issued by Merit Medical Systems, Inc.
- Company
- Merit Medical Systems, Inc
- Recall Initiated
- March 14, 2024
- Posted
- May 20, 2024
- Recall Number
- Z-1874-2024
- Quantity
- 18 units
- Firm Location
- Jordan, UT
- Official Source
- View on FDA website ↗
Reason for Recall
Their is a potential that sheaths labeled with .018 dilators may contain a .038 dilator.
Distribution
Worldwide - US Nationwide distribution in the state of CA and the countries of New Zealand and Canada.
Lot / Code Info
Lot # H2800643/ UDI: 00884450770757
Root Cause
Employee error
Action Taken
On March 18, 2024 Merit Medical issued a "Urgent Medical Device Recall Notice to affected consignees via E-Mail. Merit Medical asked consignees to take the following actions: 1. Please immediately determine if any of the devices identified in the attached Customer Response Form (CRF) are within your facility, quarantine them, and discontinue use and distribution. 2. Ensure that applicable personnel within your organization are made aware of this field action. 3. If the product has been further distributed to other facilities, institutions, or manufacturers, please ensure this notice is immediately shared with them and note the quantity distributed on the CRF. Additional distribution details may be required by health authorities. 4. Please fill out, scan and email the completed Customer Response Form to Customer Service at response@merit.com within 7 days. All affected product shipped to you must be accounted for on the CRF. 5. Please immediately return all affected lots in your possession to Merit, per the instructions in the attached CRF. 6. Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail, or by fax.