Automated Impella Controller (AIC), product number 0042-000-US with software versions V8.4 and V8.4.1.

Company
Abiomed, Inc.
Recall Initiated
October 10, 2023
Posted
December 1, 2023
Terminated
March 6, 2025
Recall Number
Z-0440-2024
Firm Location
Danvers, MA

Reason for Recall

Pump not detected as connected to controller due to software issue.

Distribution

US: AZ, CA, FL, NY & VA.

Lot / Code Info

UDI-DI (GTIN): 00813502010022, serial numbers IC3772, IC3773, IC5721 and IC4404.

Root Cause

Software design

Action Taken

An Urgent Medical Device Correction was issued vail mail October 24. 2023. Customers are advised to examine inventory for product running AIC software version: V8.4 or V8.4.1. Do not to use the product unless software is updated. Contact adminsfs@abiomed.com to schedule the software update and return the business reply form. Forward this information to relevant personnel and any facility product may have been transferred to.