Wash station: M1000 Wash Station Assembly & DITI Slide, REF 04J72-28* (*This part number is a spare to the M2000sp.)

Company
Abbott Molecular, Inc.
Recall Initiated
February 1, 2024
Posted
February 26, 2024
Recall Number
Z-1190-2024
Firm Location
Des Plaines, IL

Reason for Recall

Products that were intended for export only which were not approved or registered in accordance with U.S. specifications were shipped to U.S. customers.

Distribution

US Nationwide distribution in the states of AL, CA, IL, MA, MD, MN, NY, and PA.

Lot / Code Info

No lot number, UDI-DI 00884999006676.

Root Cause

Process control

Action Taken

The recalling firm issued letters dated and issued on 2/1/2024 via FedEx priority overnight express mail. The letter explained the background, potential impact, and necessary actions to be taken and contained an appendix listing Impacted Abbott Products and an appendix listing CE products and similar US products. The actions to be taken were to destroy all remaining inventory of the impacted products and contact their local Abbott representative for replacements. If the product has been forwarded to other laboratories, the consignee is requested to inform them of the recall and have them take the necessary actions. The letter states that laboratories should follow internal procedures in regard to a lookback of results and a lookback should be considered if results have been released that are tied to performance claims. A Customer Reply Form was enclosed to acknowledge receipt of the letter. The consignee was to record the amount of affected product that was destroyed.