Hamilton-T1, REF: MSP161339/02, ESM Shielding Set, Spare Part for ventilator

Company
Hamilton Medical Inc
Recall Initiated
July 26, 2023
Posted
September 20, 2023
Recall Number
Z-2530-2023
Quantity
4 shielding set
Firm Location
Reno, NV

Reason for Recall

Degrading capacitors on the control board of ventilators and spare parts may leak electrolyte onto the control board causing a short circuit on the board and/or the capacitor to lose function and lead to interruption of ventilation.

Distribution

Distribution US nationwide, including Puerto Rico. OUS: Not provided

Lot / Code Info

UDI: 07630002800761 / Affected Serial Numbers: 4050193, 4050210, 4050194 and 4016669

Root Cause

Nonconforming Material/Component

Action Taken

On July 26, 2023, Hamilton Medical issued Urgent Medical Device Correction via E-Mail. Hamilton Medical asked consignees to take the following actions 1. Check for affected devices in your facility. Get in contact with your local distribution partner and get the devices serviced with high priority. 2. Please sign the enclosed/attached sheet to confirm that, as an End- Customer, you have received and duly noted this Medical Device Correction.