Patient Information Center iX with Software Version Number 4.x in the following: (1) Patient Information Center iX Product Number: 866389; (2) Patient Information Center iX Expand Product Number: 866390; (3) PIC iX Essentials Product Number: 867093; (4) PIC iX Essentials Expand Product Number: 867154

This recall is currently active, issued September 19, 2023. It was issued by Philips North America Llc.

Company
Philips North America Llc
Recall Initiated
August 17, 2023
Posted
September 19, 2023
Recall Number
Z-2608-2023
Quantity
960 units
Firm Location
Cambridge, MA
Official Source
View on FDA website ↗

Reason for Recall

Patient Information Center (PIC iX) Release 4.x Surveillance Crash Caused by Intel Graphics Driver Error-with a blank screen and subsequently require a manual reboot of the hardware to restart and continue central patient monitoring. Potential for a delay in the detection of a change or deterioration in the condition of one or more patients

Distribution

Nationwide Foreign: Austria Bahrain Belgium Canada Czech Republic Denmark Finland France Germany Greece India Ireland Italy Japan Jordan Kenya Kuwait Lebanon Lithuania Netherlands Norway Poland Portugal Qatar Romania Slovakia South Africa South Korea Spain Sweden Sweden Switzerland United Kingdom Utd.Arab.Emir. Uzbekistan

Lot / Code Info

UDI: (01)00884838104594(10)4.0.1 (01)00884838104594(10)4.0.2 (01)00884838112070(10)4.1.0 Software Version Number 4.x

Root Cause

Under Investigation by firm

Action Taken

Philips issued Urgent field Safety Notice to to U.S. customers via certified mail on 8/17/23. Philips Market Organizations are responsible for distributing the letters outside of the U.S. Letter states reason for recall, health risk and action to take: If central monitoring is disrupted and blank screen displays: 1) Initiate local downtime procedures to evaluate patients at bedside and confirm local monitoring devices continue to monitor. 2) Manually reboot PIC iX to continue central monitoring This notice should be passed on to all those who need to be aware within your organization or to any organization where PIC iX 4.x devices have been transferred. 5. Actions planned by Philips to correct the problem A Philips representative will contact you to schedule a visit from a Philips Field Service Engineer who will install a software patch to your device. The patch will add defensive mitigations to force an automatic resumption of central patient monitoring without manual intervention. If central monitoring is disrupted and blank screen displays: 1) Initiate local downtime procedures to evaluate patients at bedside and confirm local monitoring devices continue to monitor. 2) Manually reboot PIC iX to continue central monitoring This notice should be passed on to all those who need to be aware within your organization or to any organization where PIC iX 4.x devices have been transferred. 5. Actions planned by Philips to correct the problem A Philips representative will contact you to schedule a visit from a Philips Field Service Engineer who will install a software patch to your device. The patch will add defensive mitigations to force an automatic resumption of central patient monitoring without manual intervention. If you need any further information, please contact your local Philips representative, or call 1-800-722-9377.