WATCHDOG Hemostasis Valve Kit, UPN H74939343A022 (outer package), UPN H74939343A020 (inner package)
- Company
- Boston Scientific Corporation
- Recall Initiated
- April 5, 2023
- Posted
- May 16, 2023
- Recall Number
- Z-1592-2023
- Quantity
- 1300 units
- Firm Location
- Maple Grove, MN
Reason for Recall
The firm received complaints of the Insertion Tool component being occluded, resulting in the inability to pass a guidewire through the tool.
Distribution
Worldwide
Lot / Code Info
GTIN 08714729996019 (outer package), GTIN 08714729965794 (inner package), Batch numbers: 30272850,30542947,30661041,30845826, 30935565
Root Cause
Process control
Action Taken
Boston Scientific issued an URGENT MEDICAL DEVICE PRODUCT REMOVAL notice on 04/05/2023 by overnight mail. The notice explained the issue and the hazard and requested the consignee segregate the devices pending return. Medical facilities who distributed the devices and distributors were directed to send the notice to their customers.