KLS Martin Battery Pack, REF: KLS BP 001, and KLS BP 040 (bulk case of 40), used in conjunction with the KLS Martin MaxDriver
This recall is currently active, issued July 5, 2023. It was issued by Pro-Dex Inc.
- Company
- Pro-Dex Inc
- Recall Initiated
- February 15, 2023
- Posted
- July 5, 2023
- Recall Number
- Z-2098-2023
- Quantity
- 2000
- Firm Location
- Irvine, CA
- Official Source
- View on FDA website ↗
Reason for Recall
Battery packs used in surgical procedures where a surgeon uses a powered screwdriver and a right-angle attachment have switched manufacturing and expiration dates on sterile pouch labels, so batteries will appear to be prematurely expired.
Distribution
US: FL
Lot / Code Info
REF: KLS-BP-001/KLS-BP-040, UDI-DI: 00862377000140, Lot: K0M6U, Expiration: 8/31/2024
Root Cause
Error in labeling
Action Taken
On 2/15/23, a recall notice was sent to a customer asking them to cease distribution and return affected product to the recalling firm. Questions can be directed to 949-769-3200.