RecallDepth

KLS Martin Battery Pack, REF: KLS BP 001, and KLS BP 040 (bulk case of 40), used in conjunction with the KLS Martin MaxDriver

This recall is currently active, issued July 5, 2023. It was issued by Pro-Dex Inc.

Company
Pro-Dex Inc
Recall Initiated
February 15, 2023
Posted
July 5, 2023
Recall Number
Z-2098-2023
Quantity
2000
Firm Location
Irvine, CA
Official Source
View on FDA website ↗

Reason for Recall

Battery packs used in surgical procedures where a surgeon uses a powered screwdriver and a right-angle attachment have switched manufacturing and expiration dates on sterile pouch labels, so batteries will appear to be prematurely expired.

Distribution

US: FL

Lot / Code Info

REF: KLS-BP-001/KLS-BP-040, UDI-DI: 00862377000140, Lot: K0M6U, Expiration: 8/31/2024

Root Cause

Error in labeling

Action Taken

On 2/15/23, a recall notice was sent to a customer asking them to cease distribution and return affected product to the recalling firm. Questions can be directed to 949-769-3200.