Teleflex (manufacturer: Vascular SOLUTIONS) D-Stat Flowable Hemostat, REF 4000; for sealing residual oozing of tissue tracts of femoral access sites that have been previously closed by suture/collagen-based hemostatic devices.
- Company
- Teleflex Llc
- Recall Initiated
- April 12, 2023
- Posted
- May 18, 2023
- Recall Number
- Z-1611-2023
- Quantity
- 1140 units
- Firm Location
- Morrisville, NC
Reason for Recall
The product contains an expired diluent vial. If the impacted product is used, the sterility of the diluent component cannot be assured as it has passed beyond the studied shelf life of the product.
Distribution
US Nationwide
Lot / Code Info
UDI/DI M20640000, Lot Numbers: 717535, exp. 04/27/2023; 717413, exp. 04/25/2023
Root Cause
Under Investigation by firm
Action Taken
Teleflex issued an URGENT - MEDICAL DEVICE RECALL notice to its consignees on 04/12/2023 via FedEx 2nd day. The notice explained the issue and the hazard and requested the following actions be taken: Medical Facilities - immediately check your inventory for product within the scope of this recall. Users should cease use and distribution of affected product and immediately quarantine the affected product. If impacted product is on hand, mark the applicable checkbox on the Acknowledgement Form and fax it to 1-855-419-8507, including Attn: Customer Service or using the email address recalls@teleflex.com. A customer service representative will contact you with a Return Goods Authorization (RGA) Number and will provide instructions for the return of products to Teleflex Medical. Distributors were directed to immediately check your inventory for product within the scope of this recall. Users should cease use and distribution of affected product and immediately quarantine the affected product. They were further directed to notify their customers, collect the response forms, and upon completion of the action forward the completed Acknowledgement Form to recalls@teleflex.com.