DePuy Mitek BIOKNOTLESS Plus w/ OC-Indicated for use in soft tissue to bone fixation in association with adequate postoperative immobilization Model Number: 212726

Company
Depuy Mitek, Inc., A Johnson & Johnson Co.
Recall Initiated
April 13, 2023
Posted
June 2, 2023
Recall Number
Z-1679-2023
Quantity
2372 units
Firm Location
Norwood, MA

Reason for Recall

IFU Warning: Applying bending force and/or bending force with impaction to the BIOKNOTLESS Anchor Inserter may cause a fragment of the inserter tip to break off during use and be retained in the patient. broken BIOKNOTLESS Anchor inserter tip detected during surgery may require removal, potentially causing bone damage and longer surgery time

Distribution

Worldwide distribution - US Nationwide and the countries of Australia, Belgium, Canada, Switzerland, China, Czech Republic, Germany, Egypt, Spain, United Kingdom, Hong Kong, Hungary, Ireland, India, Italy, Korea, Netherlands, New Zealand, Slovakia, Thailand.

Lot / Code Info

UDI-DI: 10886705002122 All lot numbers

Root Cause

Use error

Action Taken

DePuy Mitek issued Urgent Medical Device Correction (Notification) Letter on 4/13/23 via First Class Mail to US consignees. Internal and sales/distributors notified via email. International affiliates will be notified via email and will be initiated per local regulations. Letter states reason for recall, health risk and action to take: 1. Product is NOT being removed from the field and does not need to be returned. 2. Familiarize yourself with the content 2. Familiarize yourself with the content of this letter. 3. Review, complete all fields, sign, and return the attached business response form (BRF) on the last page of this letter to OneMD-Field-Actions@its.jnj.com within three (3) business days of receipt of this notification. IMPORTANT: Complete the BRF even if you do not have any of the subject product in your current inventory. 4. Forward this notice to any personnel in your facility who need to be informed. 5. If any of the subject products have been forwarded to another facility, contact that facility and provide them with this notice. 6. Post a copy of this notice in a visible area for awareness and keep a copy for your records. If you have any questions, please contact your local DePuy Synthes Sales Consultant. For Medical Information request, please visit our website: https://www.jnjmedicaldevices.com/mir.