RecallDepth

KWIK-STIK, ANA (3 Strains) QC Set, REF 5065P; Microbiological testing

Company
Microbiologics Inc
Recall Initiated
August 14, 2023
Posted
September 11, 2023
Recall Number
Z-2576-2023
Quantity
4 units
Firm Location
Saint Cloud, MN

Reason for Recall

One lot of KWIK STIK ANA (3 strains) QC Set is missing one of the strains, 6018P Parabacteroides distasonis.

Distribution

US Nationwide distribution in the states of MN, PA, TN, TX.

Lot / Code Info

UDI/DI 70845357025622, Lot Number 5065-26, Exp. 07/31/2024

Root Cause

Packaging process control

Action Taken

Microbiologics issued an Urgent Medical Device Recall notice to its consignees by email on 08/14/2023. The notice explained the problem with the device and requested the consignee either use or discard the product depending on their lab procedures and how this information affects your usage. Contact Microbiologics if replacement kit is needed.