Shanghai United Imaging Healthcare Co., Ltd. Computed Tomography X-ray System, Model:uCT 550 MD, Rx Only
This recall is currently active, issued October 21, 2025. It was issued by Uih Technologies Llc.
- Company
- Uih Technologies Llc
- Recall Initiated
- September 12, 2025
- Posted
- October 21, 2025
- Recall Number
- Z-0218-2026
- Quantity
- 22 units
- Firm Location
- Houston, TX
- Official Source
- View on FDA website ↗
Reason for Recall
Third-party outer covers can become trapped between patient table and gantry, causing mechanical interference during table movement.
Distribution
US: OUS:
Lot / Code Info
UDI:(01)06971576831036/ Serial Numbers: 353020 353021 353023 353045 353049 354001 354002 354003 354004 354005 354006 354007 354008 354009 354010 354011 354012 354013 354014 354015 354016 354017
Root Cause
Other
Action Taken
On September 12, 2025, UIH Technologies LLC, issued a recall notification to affected consignees. UIH asked consignees to take the following actions: 1. Ensure patient positioning clearance: When positioning the patient, confirm that arms and legs and clear of all table gaps and moving parts. 2. Avoid use of third-party slickers: We advise against the use of any non-UIH-approved slickers. 3. Service Contact: For technical assistance, contact your United Imaging Service engineer or reach out to our 24/7 Customer Contact Center, Toll-Free: (855) 221-1552/Direct: (832) 699-2799 4.Report adverse events: if any adverse events or quality issue occur, please report them to the FDA via the MedWatch program. 5. Please share this information with all relevant personnel at your facility, including technologist, service staff, and clinical teams.