CX Preconnect with TENACIO Pump without InhibiZone; UPNs: (1) 72404486 (15CM IP), (2) 72404488 (21CM IP);

Company
Boston Scientific Corporation
Recall Initiated
September 11, 2025
Posted
October 10, 2025
Recall Number
Z-0140-2026
Quantity
1 unit
Firm Location
Marlborough, MA

Reason for Recall

The potential for devices to experience inflation and/or deflation performance issues or difficulties.

Distribution

Worldwide distribution - US Nationwide and the country of Canada.

Lot / Code Info

1. UPN: 72404486 (15CM IP); UDI-DI: 00191506021658; Lot numbers: 1100676851, 1100676852; 2. UPN: 72404488 (21CM IP); UDI-DI: 00191506021672; Lot numbers: 1100717673;

Root Cause

Nonconforming Material/Component

Action Taken

On September 11, 2025 URGENT MEDICAL DEVICE RECALL letters were sent to customers. Actions to be taken: For implanted devices: If the device is working properly, no action is required. Further evaluation is recommended for patients who were implanted with an affected device AND are having difficulty inflating or deflating the device. - Check inventory for affected product and return any unused devices. Immediately cease further distribution or use of any unused affected devices. Please pass this notice to any healthcare professional from your organization that needs to be aware and to any organization where the potentially affected devices have been transferred. Please provide Boston Scientific with details of any affected devices that have been transferred to another organization.