IDM-MICRO-R. Electrical wheelchair component.

This recall is currently active, issued September 2, 2025. It was issued by Mo-Vis Bvba Biebuyckstraat 15D Nevele Belgium.

Company
Mo-Vis Bvba Biebuyckstraat 15D Nevele Belgium
Recall Initiated
July 10, 2025
Posted
September 2, 2025
Recall Number
Z-2456-2025
Quantity
15 units
Official Source
View on FDA website ↗

Reason for Recall

Devices with firmware versions 2.3 and lower may experience timing variations between software tasks, which could result in skipping the necessary neutral check and the wheelchair producing unintended movement upon power-up when the joystick is out of neutral.

Distribution

US Nationwide distribution in the state of TX.

Lot / Code Info

Model No. PRSPS0014; UDI-DI: 05407008320836; Serial No. 1000 to 1690.

Root Cause

Device Design

Action Taken

An URGENT: MEDICAL DEVICE RECALL R-NET JOYSTICKS notification dated 07/10/2025 was sent to consignees instructing them to upgrade affected devices according to the provided Annex 2 or by emailing contact@mo-vis.com to coordinate return. The software upgrade of affected devices should be conducted during the next scheduled annual maintenance of the joystick. These actions should be verified. Questions can be directed via email to contact@mo-vis.com or by phone at +32 9 335 28 60.