RecallDepth

Sparrow Ascent Patient Controller, REF's: 11-110-K, 11-310-K, 11-320-K, 11-321-K, 11-310-K-28, 11-320-K-28, 11-321-K-28

This recall is currently active, issued August 11, 2025. It was issued by Spark Biomedical Inc.

Company
Spark Biomedical Inc
Recall Initiated
July 2, 2025
Posted
August 11, 2025
Recall Number
Z-2287-2025
Quantity
78 units
Firm Location
Richmond, TX
Official Source
View on FDA website ↗

Reason for Recall

Due to manufacturing issue, neurostimulator device may experience "cable disconnect" error message and interruption to stimulation output.

Distribution

US Nationwide distribution.

Lot / Code Info

Lot numbers: V2408xxxx and runs through V2519xxxx UDI: 11-110/00850052017033; 11-310-K/00860005396948; 11-320-K/00860005396986; 11-321-K/00850052017163; 11-310-K-28/00850052017088; 11-320-K-28/00850052017071; 11-321-K-28/00850052017170

Root Cause

Process control

Action Taken

On July 2, 2025, Spark Biomedical initiated a Urgent Medical Device Recall via phone. During the phone conversation, Spark Biomedical will ask consignees to take the following actions: 1. Spark Biomedical will replace all affected00 units at no cost. Even if the units appear to be working. 2. Current devices may continue to be used until they are replaced. 3. Please perform a check of the cable to patient controller connection before use. If this connection is loose or moves, please discontinue use of that patient controller device and quarantine that device until the replacement arrives. 4. Please notify all appropriate personnel at your facility that need to know about this recall.