REF: 666289 , BD FACS, 7-Color Setup Beads, 25 Runs, CE, IVD. Used in flow cytometry for fluorescence compensation.

Company
Becton, Dickinson And Company, Bd Bio Sciences
Recall Initiated
June 3, 2025
Posted
July 3, 2025
Recall Number
Z-2079-2025
Quantity
278 units
Firm Location
Milpitas, CA

Reason for Recall

Due to customer complaints, errors in spectral overlap factors table for 7-color setup beads may affect accurate setup of the flow cytometry instrument prior to analyzing samples.

Distribution

US: Arkansas, Arizona, California, Colorado, Connecticut, District of Columbia, Delaware, Florida, Georgia, Iowa, Illinois, Indiana, Kansas, Kentucky, Louisiana, Massachusetts, Maryland, Michigan, Minnesota, North Carolina, Nebraska, New Jersey, Nevada, New York, Ohio, Oklahoma, Oregon, Pennsylvania, Puerto Rico, Rhode Island, South Dakota, Tennessee, Texas, Utah, Virginia, Wisconsin, West Virginia Argentina, Australia, Belgium, Brazil, Canada, Chile, China, Colombia, Dominican Republic, Ecuador, Hong Kong, India, Japan, South Korea, Mexico, Malaysia, Nepal, New Zealand, Peru, Pakistan, Singapore, El Salvador, Taiwan OUS:

Lot / Code Info

Lot # 85637 & 85638/ UDI: 382906662897

Root Cause

Process control

Action Taken

On June 3, 2025, Becton, Dickinson issued a "Urgent: Medical Device Recall Correction" Notification to affected customers via mail and E-Mail. BD asked consignees to take the following action: 1: Immediately inspect your inventory for the specific catalog and lot numbers listed. It is recommended that the clinical laboratory staff review and segregate the affected lots. 2. Remove and discard prior affected label insert with incorrect spectral overlap factor values included on kit box prior to use. 3. The wrong SOF (Spectral Over Factors) values can be corrected by entering the right SOF values provided in Table 1. 4. Ensure the right SOF values are entered in the flow cytometry instrument prior to sample acquisition. 5. Customers who have not used the affected product may proceed to enter the SOF values. 6. If the affected products have been fully or partially used and subjects' samples have been processed and acquired on an instrument with the incorrect SOF values, please review the results. 7. If the subject sample(s) is not available anymore, this may require the subject to return to collect an additional sample for testing purposes. 8. Ensure the contents of this Product Recall Notification are read and understood. 9. Share and post this customer letter with the users of this application at your facility and forward to any customers to whom you may have distributed the product to ensure awareness.