RecallDepth

Catalog No. 8888160119; Umbilical Vessel Catheter Insertion Tray with 2.5 and 5 French UVCs

Company
Cardinal Health 200, Llc
Recall Initiated
May 2, 2025
Posted
May 29, 2025
Recall Number
Z-1870-2025
Quantity
370
Firm Location
Waukegan, IL

Reason for Recall

Packaging defect may compromise sterility of the product. Use of a non-sterile product on the patient could lead to infection in neonates.

Distribution

Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, IL, IN, KS, LA, MA, MD, ME, MI, MN, MO, MS, NC, NJ, NM, NY, OH, OK, OR, PA, PR, RI, SD, TN, TX, UT, VA, WA, WI, WV and the countries of Brazil, Canada, Chile, Colombia, Singapore, South Africa.

Lot / Code Info

UDI-DI (ea) 10192253040289; UDI-DI (box) 50192253040287 Lot 2502400146

Root Cause

Package design/selection

Action Taken

The firm started notifying customers on May 2, 2025 via Urgent Medical Device Product Recall letters. Customers were instructed to remove any devices currently in use. In addition, customers should review their inventory and segregate/quarantine all affected product. If product has been further distributed, the distributor should notify all downstream customers. Customers are asked to return an acknowledgement form whether they have affected product or not. Customers should contact the firm to arrange for return and credit or questions regarding suitable alternative products. For questions related to this recall, please contact the market action team at: GMB-FieldCorrectiveAction@cardinalhealth.com or call 800-292-9332.