RecallDepth

Brand Name: Contour Product Name: Contour Gold Eyelid Weight Model/Catalog Number: LL3012 Product Description: Contour Gold Eyelid Weight, 1.2 gm MedDev Contour and ThinProfile Eyelid Weight Implants are designed for the gravity-assisted treatment of protracted permanent lagophthalmos, usually resulting from facial paralysis.

This recall is currently active, issued May 9, 2025. It was issued by Meddev Corp.

Company
Meddev Corp
Recall Initiated
July 12, 2024
Posted
May 9, 2025
Recall Number
Z-1745-2025
Quantity
34 weights
Firm Location
Sunnyvale, CA
Official Source
View on FDA website ↗

Reason for Recall

Due to mislabeling of products

Distribution

U.S. Nationwide distribution in the states of CA, CO, CT, FL, IA, IL, MA, ME, MI, NE, NY, OH, OK, OR, SC, TX, and WV.

Lot / Code Info

Model No LL3012; UDI code: 010081314202039917260727107228 Lot Number: 7228

Root Cause

Process control

Action Taken

On 07/12/2024, the firm sent a customer notification letter via first class mail to inform customers that the labels on the Contour Gold Eyelid Implants LL3010/Lot # 7227 potentially do not match the weight; the implants labeled LL3010 (1.0 gm) may contain LL3012 (1.2 gm). Customers are instructed to check their inventory for any/all of the affected products and to contact MedDev Corporation to request for an RGA number in or return affected products and receive replacement products. For questions/concerns contact Quality Compliance Specialist at 1-800-543-2789 or email catherine@meddev-corp.com An updated "URGENT: MEDICAL DEVICE RECALL" Letter was distributed to customers on 04/17/2025. Updated Customer Letter is informing customer that the potential product mix-up may include an additional product and lot number (Model Number LL3012 and Lot Number 7228)

More recalls by Meddev Corp

View all recalls by this company →