ActiveAid 922, Shower/Commode Chair

This recall has been terminated (originally issued January 4, 2023).

Company
Altimate Medical, Inc.
Recall Initiated
November 30, 2022
Posted
January 4, 2023
Terminated
August 31, 2023
Recall Number
Z-0880-2023
Quantity
1 unit
Firm Location
Morton, MN
Official Source
View on FDA website ↗

Reason for Recall

Device manufactured with a component with affects the foldability of the frame, not allowing it to be secured by the frame strap in the folded position.

Distribution

US Nationwide distribution in the state of Texas.

Lot / Code Info

Serial Number 90071504001

Root Cause

Component change control

Action Taken

Altimate Medical notified its sole consignee by telephone on 11/30/2022 and followed with an Urgent Medical Device Recall letter on 12/06/2022. The notice stated the problem and requested the following action: "- Acknowledge receipt of this notification. Complete and return the attached Acknowledgement form. - Inform the customer not to use the chair if this has not already done. - Provide replacement chair 90071504001 under RMA000737 to AMI - Inform Altimate when the replacement chair is provided to the customer. Complete and return the attached Verification form."