Philips Laser System, REF LAS-100, Intermittent Operation, Duty cycle dependent upon device in use 100 - 240 VAC, 50/60HZ, 16 AM

Company
Spectranetics Corporation
Recall Initiated
June 1, 2022
Recall Number
Z-1351-2022
Quantity
105 units
Firm Location
Colorado Springs, CO

Reason for Recall

The LAS-100 Laser system may detect an inoperable hardware component during power up, which results in an error code and the system not being operable until code is cleared.

Distribution

US: AR AZ CA CO CT FL GA HI IA IL IN KS LA MA MD ME MI MN MO NC NJ NY OK PA TN TX VA OUS: Saudi Arabia, Denmark, France, Greece, India, Italy, Netherlands, Serbia, Spain, Sweden, United Kingdom

Lot / Code Info

Serial Numbers: 100000 to 100104 UDI/DI: (01)00813132027452

Root Cause

Device Design

Action Taken

On June 08, 2022 Spectranetics issued a "Urgent Medical Device Correction" notification to affected consignees via USPS. The recall notification asked consignees to take the following actions: 1. Philips internal testing confirms that 1-3 reboots of the system will temporarily resolve the issue. Philips also recommends the continued use of the Philips Laser System, and to follow the Operator s Manual, as there is no expected harm to the user or patient. Philips recommends notifying all Philips Laser System users within your facility of this communication and retain a copy of this letter for reference. To acknowledge receipt of this notification, please complete, sign, and return the Customer Reply Form, included with this letter, within 30 days upon receipt of this notice to the following Email: igtdc.r@philips.com If you are a distributor or have forwarded affected device(s) to another end user, it is imperative that all end-users with affected devices receive this URGENT Medical Device Correction. Therefore, send a copy of this notification to any customer to whom you have distributed the affected product. If you need any further information or support concerning this issue, please contact your local Philips representative or Philips IGTD Customer Service: Philips Laser System (PLS) Customer Service: Phone: 1-800-231-0978 Email: laserserviceops@philips.com Hours of Operation: Monday - Friday 9:00AM 5:00PM MST

More recalls by Spectranetics Corporation

View all recalls by this company →