RecallDepth

AESCULAP S4 Cervical Cross Connector L 42-58mm, S4C CROSS CONNECTOR VARIABLE 42-58MM, product code SW117T. Used in spinal fusion surgery.

Company
Aesculap Implant Systems Llc
Recall Initiated
September 8, 2020
Terminated
April 10, 2024
Recall Number
Z-1448-2022
Quantity
27
Firm Location
Center Valley, PA

Reason for Recall

Insufficient clamping force of the cross connectors in the spinal surgical system which may lead to movement of the connectors on the rods and/or microparticle abrasion.

Distribution

United States Nationwide distribution in the states of CA, CO, IL, IN, MI, NC, NV, OH, PA, SC, TX & WY.

Lot / Code Info

UDI-DI: 04038653362349

Root Cause

Device Design

Action Taken

An Urgent Medical Device Removal Notification issued September 08, 2020 sent via USPS certified mail. The notice identifies the affected part numbers S4C CROSS CONNECTOR SW112T, SW113T, SW114T, SW115T, SW116T and SW117T and asks distributors and customers to identify existing inventory, quarantine, and return affected inventory. If further distributed, identify and notify customers. The notice includes an acknowledgement form to be completed and returned to Aesculap Inc. via email at qa-recalls@aesculapusa.com. Questions are to be directed to Lindsay Chromiak, Manager, QA Compliance and Vigilance (610) 984-9072.

More recalls by Aesculap Implant Systems Llc

View all recalls by this company →