Centricity PACS-IW with Universal Viewer - Product Usage: Is a device that displays medical images (including mammograms) and data from various imaging sources.

Company
Ge Healthcare, Llc
Recall Initiated
August 6, 2021
Posted
September 22, 2021
Recall Number
Z-2502-2021
Quantity
392 (219 US; 173 OUS) all products
Firm Location
Waukesha, WI

Reason for Recall

Image acquisition failures and synchronization failure with the Centricity Enterprise Archive

Distribution

AK, AZ, CA, CO, DE, FL, GA, IA, IN, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NJ, NM, NV, NY, OH, OK, OR, PA, SD, TX, UT, WA, WI, WV and WY. Austria, Brazil, Canada, France, Germany, Ghana, India, Ireland, Israel, Italy, Kuwait, Netherlands, Panama, Poland, Russia, Saudi Arabia, Slovenia, South Africa, Spain, Switzerland, Turkey, United Kingdom and Uruguay.

Lot / Code Info

2068177-001 (Centricity PACS-IW with Universal Viewer software versions 5.0.x with PACS-IW foundation) System ID: 083055248481416, 1-121967267, 1-429329281, DGC19607785, HC3062UV01, HCIT1401012UW, HCIT4331279UV, PACS-IW-127, PACS-IW-219, PACS-IW-2770, PACS-IW-323, PACS-IW-421, RICHMONDU14767331, RICHMONDU15531177N1, X1347906 / X1347909, X1347906 / X1347909, YP315021, ZA2412PAC21 and ZA2533PAC1

Root Cause

Software design

Action Taken

This issue will be resolved by sending an Urgent Medical Device Correction letter to customers with affected software versions on August 06, 2021. This letter informs customers of the issues and provides safety instructions. GE Healthcare will correct the issue through a product change. If you have any questions or concerns regarding this notification, please contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.