Critical Care Ventilator, Catalog Number(s): 8422200: Babylog VN600 - Product Usage: used for treating patients who require temporary or longer term respiratory support for different medical reasons.

Company
Draeger Medical, Inc.
Recall Initiated
April 16, 2021
Terminated
February 6, 2024
Recall Number
Z-1692-2021
Quantity
2 units
Firm Location
Telford, PA

Reason for Recall

Three separate and unrelated problems attributed to the software used in the Evita V600, Evita V800, Babylog VN600 and Babylog VN800: 1. Restart of ventilation unit. 2. Incorrect FiO2 high and FiO2 low alarms. 3. SmartCare/PS (SC/PS) suspends weaning

Distribution

US Nationwide distribution in the states of PA, TX, MD, FL, TX, WA, MS, NY.

Lot / Code Info

All lots/serial numbers manufactured from 1/1/2019-3/31201

Root Cause

Software design

Action Taken

In April, 2021, Draeger issued an Urgent Medical Device Correction notice to customer via letter notifying them that Draeger became aware of three separate and unrelated problems attributed to the software used in the Evita V600, Evita V800, Babylog VN600 and Babylog VN800: