RecallDepth

OSTEOVATION 5CC IMPACTFORMULA. REF/UDI: 390-2105/00813845020085 - Product Usage: The product provides a bone void filler that is replaced by bone during the healing process.

Company
Osteomed, Llc
Recall Initiated
April 2, 2020
Posted
May 29, 2020
Terminated
May 5, 2021
Recall Number
Z-2188-2020
Quantity
1724
Firm Location
Addison, TX

Reason for Recall

Incorrect shelf-life for bone void filler kits; the product may not maintain its setting characteristics for the labeled shelf-life duration.

Distribution

Worldwide distribution - US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA. MA, MD, ME. MI, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV and WY and the countries of AUSTRALIA, ITALY, KOREA, MALAYSIA, PERU, SINGAPORE and THAILAND.

Lot / Code Info

LOT 17060509, 17102710, 17040533, 17051911, 1126986, 1128789, 1134232, 1136330, 1144043 and 1142210

Root Cause

Vendor change control

Action Taken

On 04/02/2020, an URGENT Medical Device Recall letter was sent to all consignees affected by the recall. OsteoMed has requested that affected customers: (1) review this letter in full, review current inventory, and complete and return the Acknowledgement and Receipt Form via email or fax; and (2) return unused inventory via Federal Express (using OsteMed account)

More recalls by Osteomed, Llc

View all recalls by this company →