Heartware HVAD Pump Implant Kit, Product (REF) Number 1104JP
This recall is currently active, issued April 3, 2020. It was issued by Heartware, Inc..
- Company
- Heartware, Inc.
- Recall Initiated
- April 3, 2020
- Recall Number
- Z-2057-2020
- Quantity
- 85 units
- Firm Location
- Miami Lakes, FL
- Official Source
- View on FDA website ↗
Reason for Recall
The Outflow Graft may be subject to tears and the Strain Relief screw may break during the pre-implant pump assembly and attachment to the HVAD Pump.
Distribution
Worldwide Distribution
Lot / Code Info
GTIN 00888707000697, Serial numbers: HW34700, HW34703, HW34710, HW34713 GTIN 00888707005838, Serial Numbers: HW40383, HW40412, HW40413, HW40414, HW40415, HW40417, HW40418, HW40419, HW40420, HW40421, HW40422, HW40423, HW40424, HW40425, HW40426, HW40428, HW40429, HW40431, HW40465, HW40471, HW40472, HW40473, HW40474, HW40475, HW40476, HW40477, HW40478, HW40479, HW40480, HW40552, HW40553, HW40554, HW40555, HW40556, HW40557, HW40564, HW40565, HW40580, HW40599, HW40600, HW40602, HW40603, HW40604, HW40607, HW40609, HW40611, HW40612, HW40613, HW40649, HW40650, HW40651, HW40652, HW40653, HW40655, HW40656, HW40657, HW40658, HW40659, HW40660, HW40661, HW40662, HW40663, HW40664, HW40666, HW40667, HW40668, HW40789, HW40816, HW40881, HW40950, HW40951, HW40952, HW40953, HW40954, HW40955, HW40956, HW40957, HW40958, HW40960, HW40961, HW41045
Root Cause
Device Design
Action Taken
The firm disseminated a Medical Device Safety Alert by letter beginning on 04/03/2020. The notice explained he potential for damage that can occur during the procedure. Specifically, the Outflow Graft may be subject to tears and the Strain Relief screw may break during the pre-implant pump assembly and attachment to the HVAD Pump. The letter also provided additional steps for assembly and attachment to reduce the risk of damage and tearing during the assembly procedure which was provided in Appendix A.