Heartware HVAD Pump Implant Kit, Product (REF) Number 1104JP

Company
Heartware, Inc.
Recall Initiated
April 3, 2020
Recall Number
Z-2057-2020
Quantity
85 units
Firm Location
Miami Lakes, FL

Reason for Recall

The Outflow Graft may be subject to tears and the Strain Relief screw may break during the pre-implant pump assembly and attachment to the HVAD Pump.

Distribution

Worldwide Distribution

Lot / Code Info

GTIN 00888707000697, Serial numbers: HW34700, HW34703, HW34710, HW34713  GTIN 00888707005838, Serial Numbers: HW40383, HW40412, HW40413, HW40414, HW40415, HW40417, HW40418, HW40419, HW40420, HW40421, HW40422, HW40423, HW40424, HW40425, HW40426, HW40428, HW40429, HW40431, HW40465, HW40471, HW40472, HW40473, HW40474, HW40475, HW40476, HW40477, HW40478, HW40479, HW40480, HW40552, HW40553, HW40554, HW40555, HW40556, HW40557, HW40564, HW40565, HW40580, HW40599, HW40600, HW40602, HW40603, HW40604, HW40607, HW40609, HW40611, HW40612, HW40613, HW40649, HW40650, HW40651, HW40652, HW40653, HW40655, HW40656, HW40657, HW40658, HW40659, HW40660, HW40661, HW40662, HW40663, HW40664, HW40666, HW40667, HW40668, HW40789, HW40816, HW40881, HW40950, HW40951, HW40952, HW40953, HW40954, HW40955, HW40956, HW40957, HW40958, HW40960, HW40961, HW41045

Root Cause

Device Design

Action Taken

The firm disseminated a Medical Device Safety Alert by letter beginning on 04/03/2020. The notice explained he potential for damage that can occur during the procedure. Specifically, the Outflow Graft may be subject to tears and the Strain Relief screw may break during the pre-implant pump assembly and attachment to the HVAD Pump. The letter also provided additional steps for assembly and attachment to reduce the risk of damage and tearing during the assembly procedure which was provided in Appendix A.

More recalls by Heartware, Inc.

View all recalls by this company →