Omnipod 5 Automated Insulin Delivery System Product catalog numbers: PT-000409: ASM, Omnipod 5, PDM, Programmed PT-000428: Adaptor, Charging, Noetic Nuu N5004L, Horizon PDM, US PT-000429: Cable, Charging, Omnipod 5, Horizon PDM

Recall Initiated
November 14, 2022
Posted
December 7, 2022
Recall Number
Z-0423-2023
Quantity
129,086
Firm Location
Acton, MA

Reason for Recall

The firm has become aware of an issue with the Omnipod 5 Controller where the Controller charging port and charging cable are melting, deforming, or discoloring due to heat generated by increased resistance in the Controller port. The excess heat may cause minor burns if touched, and the Omnipod 5 Controller may fail to charge, requiring the user to switch to their backup insulin plan.

Distribution

Domestic distribution nationwide.

Lot / Code Info

ALL serial numbers/ controller lots are currently in scope of this recall

Root Cause

Device Design

Action Taken

On November 14, 2022, the firm initiated their Press Release at https://investor.insulet.com/news-releases/news-release-details/insulet-issues-nationwide-voluntary-medical-device-correction and notified affected customers through Urgent Medical Device Correction letters. Customers were instructed to monitor their Omnipod 5 Controller for the following issues: - Charging port or charging cable appears melted, deformed, or discolored - Controller experiences extreme overheating (uncomfortable to hold) or emits an odor while charging If customers experience any of the issues above, they should disconnect the charger from the power outlet and do NOT charge the Omnipod 5 Controller. Promptly contact Insulet s dedicated Customer Care team specifically trained to help you at 1-800-641-2049 to request a replacement device. If you have a compatible Android smartphone (please see omnipod.com/compatibility), you may want to switch to using the Omnipod 5 App on your smartphone. Customers were provided with additional precautions to take to continue use of the device.