Zio AT Clinical Reference Manual, Part: ALB0031, and Zio AT Important Information Pamphlet, Part: ALB0034. For Zio Ambulatory ECG monitoring system

Company
Irhythm Technologies, Inc.
Recall Initiated
September 28, 2022
Posted
November 4, 2022
Recall Number
Z-0183-2023
Quantity
1,443
Firm Location
Cypress, CA

Reason for Recall

Ambulatory ECG monitoring system labeling update: 1) Clinically actionable arrhythmia notifications delayed until patient registration is complete; 2) Patches limited to 100 symptomatic and 500 asymptomatic wireless transmissions; once the maximum transmission limit is reached for either type, any further transmissions for that type will cease. Risk of delayed reporting to health providers.

Distribution

US: MD, NY, FL, TN, WA, AL, MA, AZ, CA, KS, HI, PA, NJ, TX, MN, VA, AR, OR, NC, CO, KY, MS, LA, MI, WI, WV, IN, ID, SC, CT, SD, IA, NV, OK, OH, IL, VT, NH, NE, NM, MO, ME, DC, RI, MT, DE, GA, ND

Lot / Code Info

UDI-DI: 00869770000210; Model/Revision: ALB0031/07, ALB0034/03

Root Cause

Device Design

Action Taken

On 9/28/2022, Field Safety Notice Labeling Corrections were sent to customers who were informed of labeling changes and that the firm intends to enhance the process of notifying patients that they are approaching the maximum limit for either symptomatic or asymptomatic transmissions by auto shipping another Zio AT patch to the patient. Customers are asked to do the following: 1) Please review this letter and disseminate it to the appropriate system providers and any personnel at your facility. Please share this notification with all device users within your facility to ensure they are aware of this notice. 2) Complete and return the customer reply form. Further questions can be directed to customer care at (888) 693-2401 or support@irhythmtech.com or your local sales representative.