Product/Part: YelloPort Elite 12x75 Pencil Point Trocar/ET1207503; YelloPort Elite Hasson Adaptor 10mm/EA10NH. For use in laparoscopic procedures.

Company
Surgical Innovations Ltd 6 Clayton Wood Bank Lds Leeds United Kingdom
Recall Initiated
January 20, 2022
Posted
September 29, 2022
Recall Number
Z-1852-2022
Quantity
19

Reason for Recall

Laparoscopic devices have a sterile symbol included on labels but devices are non-sterile.

Distribution

US: MA

Lot / Code Info

Part/UDI-DI/Lot: ET1207503/05051986012438/735479, EA10NH/05051986012100/734029

Root Cause

Process control

Action Taken

On 01/20/2022, Urgent Field Safety Notices were emailed to customers who were asked to do the following: 1) Quarantine affected devices and return to the recalling firm. 2) If the devices are no longer in stock, confirm whether devices were used clinically, and if so, whether the devices were sterilized in accordance with the Instructions for Use prior to use. 3) Pass on this notice to all those who need to be aware within your organization or to any organization where the potentially affected devices were transferred. 4) Complete response form. The recalling firm provided the following email: quality@surginno.co.uk