RecallDepth

RayStation 4-11B, and RayPlan 1, 2, 7-11B including some service packs: RayStation 4.0 (4.0.0.14) to RayStation 5 Service Pack 2 (5.0.2.35), RayStation 5 Service Pack 3 (5.0.3.17), RayStation 6/RayPlan 2 (6.0.0.24), RayStation 6/RayPlan 2 Service Pack 1 (6.1.1.2), RayStation 6/RayPlan 2 Service Pack 2 (6.2.0.7), RayStation 6/RayPlan 2 Service Pack 3 (6.3.0.6), RayStation/RayPlan 7 (7.0.0.19), RayStation/RayPlan 8A (8.0.0.61), RayStation/RayPlan 8A Service Pack 1 (8.0.1.10) Software versions: 4.0.0.14, 4.0.3.4, 4.5.1.14, 4.7.2.5, 4.7.3.13, 4.7.4.4, 4.7.5.4, 4.7.6.7, 4.9.0.42, 5.0.1.11, 5.0.2.35, 5.0.3.17, 6.0.0.24, 6.1.1.2, 6.2.0.7, 6.3.0.6, 7.0.0.19, 8.0.0.61, 8.0.1.10, 8.1.0.47, 8.1.1.8, 8.1.2.5, 9.0.0.113, 9.1.0.933, 9.2.0.483, 10.0.0.1154, 10.0.1.52, 10.1.0.613, 10.1.1.54, 11.0.0.951, 11.0.1.29, 11.0.3.116 or 12.0.0.932

Company
Raysearch America Inc
Recall Initiated
February 11, 2022
Recall Number
Z-0850-2022
Quantity
653 (US 149, OUS 504)
Firm Location
York, NY

Reason for Recall

Incorrect source to skin or surface distance (SSD) calculation may result in dose error when using RayStation system.

Distribution

US Nationwide distribution in the states of AK, AL, AR, AZ, CA, CT, DC, DE, FL, GA, HI, IA, ID, IL, KS, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NH, NJ, NM, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI.

Lot / Code Info

UDI # 07350002010020, 07350002010013, 07350002010082, 07350002010075, 07350002010242, 07350002010068, 07350002010112, 07350002010136, 07350002010129, 07350002010204, 07350002010235, 07350002010174, 07350002010266, 07350002010297, 07350002010303, 07350002010365, 07350002010310, 07350002010471, 07350002010389, 07350002010433, 07350002010440, 07350002010426

Root Cause

Under Investigation by firm

Action Taken

A field safety notice dated February 07, 2022 was issued via email on February 11, 2022 which includes work around instructions to make sure that the SSD value is correct. Identify affected product via software version number(s). Notify all users and acknowledge receipt of the notice. For more information contact quality@raysearchlabs.com.

More recalls by Raysearch America Inc