MUSE Cardiology Information System, Model Numbers: a) 3101011-040, b) 3101010-401, c) 3101010-009, d) 3101010-008

Company
Ge Healthcare, Llc
Recall Initiated
March 16, 2022
Recall Number
Z-0919-2022
Quantity
147 units
Firm Location
Waukesha, WI

Reason for Recall

Two scenarios may cause edits to measurements and diagnosis statements can be lost after a test is Signed in the MUSE NX web client.

Distribution

US and Australia, Austria, Bahrain, Belgium, Brazil, Canada, Denmark, England, France, Germany, Ireland, Italy, Korea, Malaysia, Netherlands, Norway, Poland, Qatar, Saudi Arabia, Singapore, Spain, Sweden, Taiwan, Turkey

Lot / Code Info

a) 3101011-040, UDI (01)00840682142809(10)MUSENXSP2, Serial Numbers: SSR20435644TA, SSR20085352TA, SSR19455267TA, SSR20245386TA, SSR20225354TA, SSR20235322TA, SSR19485286TA, SSR20265441TA, SSR20435653TA, SSR20235349TA, SSR20315513TA, SSR20045337TA, SSR19415230TA, SSR19425238TA; UDI (01)00840682142809(10)MUSENXSP3, Serial Numbers: SSR19305164TA, SSR21255865TA, SSR20525351TA, SSR20525347TA, SSR20325524TA, SSR21326015TA, SSR21025368TA, SSR21065487TA, SSR19395224TA; UDI (01)00840682142809(10)MUSENXSP4, Serial Numbers: SSR19445252TA, SSR21035400TA, SSR21536318TA, SSR22056438TA, SSR20225343TA, SSR20515315TA, SSR20265429TA, SSR20465706TA, SSR20315501TA  b) 3101010-401, UDI (01)00840682142809(10)MUSENXSP2, Serial Numbers: SSR19305178TA, SSR19265115TA, SSR20315505TA, SSR20405600TA, SSR20415608TA, SSR20255413TA, SSR20385590TA, SSR20265439TA, SSR19485282TA, SSR19435251TA, SSR20235360TA, SSR20125313TA, SSR20325529TA, SSR20125350TA, SSR19445257TA, SSR20035336TA, SSR19515306TA, SSR19415228TA, SSR20235369TA, SSR20355576TA, SSR20515344TA, SSR20225349TA, SSR20415610TA, SSR20335533TA, SSR20235365TA, SSR20425623TA, SSR19525317TA, SSR20235346TA; UDI (01)00840682142809(10)MUSENXSP3, Serial Numbers: SSR21185714TA, SSR19285150TA, SSR21085519TA, SSR21356058TA, SSR20455682TA, SSR21215784TA, SSR21275895TA, SSR21285908TA, SSR21205767TA, SSR20295480TA, SSR21045424TA, SSR21175691TA, SSR21055448TA, SSR19355213TA, SSR21065475TA, SSR21055459TA, SSR21165669TA, SSR21185705TA, SSR21195741TA, SSR21125596TA, SSR21235844TA, SSR20465707TA, SSR21315999TA, SSR21426117TA, SSR21155657TA, SSR21426114TA, SSR21035404TA, SSR21295941TA, SSR20235345TA, SSR21055456TA, SSR20315499TA, SSR20315500TA, SSR20275464TA, SSR21185719TA, SSR21185720TA, SSR21386052TA, SSR21386053TA, SSR21386054TA, SSR21436118TA, SSR21436122TA, SSR21436123TA, SSR20235325TA, SSR21185731TA, SSR20315494TA, SSR20315493TA, SSR20265446TA, SSR21356057TA; UDI (01)00840682142809(10)MUSENXSP4, Serial Numbers: SSR21476197TA, SSR21305971TA, SSR21506267TA, SSR21185730TA, SSR21396072TA, SSR21486216TA, SSR21476205TA, SSR21356047TA, SSR21215780TA, SSR21506263TA, SSR21225815TA, SSR21496242TA, SSR21526307TA, SSR19505289TA, SSR21516287TA, SSR22046398TA, SSR21516281TA, SSR21526299TA, SSR21416095TA, SSR21235846TA, SSR21336022TA, SSR20295484TA, SSR21275889TA, SSR21496247TA, SSR22096508TA, SSR21506252TA, SSR21466153TA, SSR22066453TA;  c) 3101010-009, UDI (01)00840682142809(10)MUSENXSP3, Serial Numbers: SSR21095569TA, SSR21135614TA, SSR21025386TA; UDI (01)00840682142809(10)MUSENXSP4, Serial Numbers: SSR22026369TA, SSR22046388TA;  d) 3101010-008, UDI (01)00840682142809(10)MUSENXSP2, Serial Numbers: SSR20235354TA, SSR20505775TA, SSR20255408TA; UDI (01)00840682142809(10)MUSENXSP3, Serial Numbers: SSR21366065TA, SSR21065468TA, SSR21316007TA; UDI (01)00840682142809(10)MUSENXSP4, Serial Numbers: SSR21486211TA

Root Cause

Software Manufacturing/Software Deployment

Action Taken

GE Healthcare notified its consignees by letter titled URGENT MEDICAL DEVICE CORRECTION on March 16, 2022. The letter explained the two scenarios which would produce the loss of edits and provided recommendations and a workaround. The firm will correct all affected products at no cost to you. The firm will be issuing a software patch. A GE Healthcare representative will contact affected customers to arrange for the correction. If you have any questions or concerns regarding this notification, contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative