RecallDepth

Discovery NM670 Pro Model # H3100XB

This recall has been terminated (originally issued August 14, 2020).

Company
Ge Healthcare, Llc
Recall Initiated
August 14, 2020
Terminated
May 6, 2021
Recall Number
Z-2997-2020
Quantity
9 units
Firm Location
Waukesha, WI
Official Source
View on FDA website ↗

Reason for Recall

Rotor bearing screws were found loose on detector.

Distribution

USA : AR CT FL GA IL KS KY LA MA MD MI MN MO NE NJ NV NY OH OK RI SC TN TX VA VT WA WI OUS: ARGENTINA AUSTRIA BELGIUM BRAZIL Canada CHINA FRANCE GERMANY GREECE INDIA ISRAEL ITALY JAPAN Korea (Republic of) NETHERLANDS NORWAY PANAMA POLAND PORTUGAL Russia SPAIN SWEDEN SWITZERLAND TAIWAN THAILAND TURKEY UNITED KINGDOM VIETNAM

Lot / Code Info

Serial # PRGX52001 System ID: NMPRGX52001 Serial # PRGX52003 System ID: 082426210033 Serial # PRGX52008 System ID: 082426120037 Serial # PRGX52005 System ID: 082426120039 Serial # PRGX52007 System ID: 082426290010 Serial # PRGX52004 System ID: 082426180017 Serial # PRGX52009 System ID: 082426130053 Serial # PRGX52002 System ID:082426150044 Serial # PRGX52006 System ID:082426160065

Root Cause

Use error

Action Taken

GE Healthcare issued an Urgent Medical device correction notification on August 14, 2020. It states the following: 1. GE Healthcare will inspect and, if required, correct all affected products at no cost to you. 2. A GE Healthcare representative will contact you to arrange for the correction 3. Please complete the acknowledgement form and return it to GE Healthcare promptly upon receipt and no later than 30 days from receipt to nm.fmi40890.responses@ge.com If you have any questions or concerns regarding this notification, please contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.

More recalls by Ge Healthcare, Llc

View all recalls by this company →