RecallDepth

Stryker T2 Alpha Delta Strike Plate IMN Instruments - part of the IMN Instruments System and is used to support insertion and withdrawal of nails through the Targeting Device Catalog Number: 2351-0050

Company
Stryker Gmbh Bohnackerweg 1 Selzach Switzerland
Recall Initiated
December 5, 2019
Terminated
October 9, 2020
Recall Number
Z-0780-2020
Quantity
433 (289 units U.S. / 144 units OUS)

Reason for Recall

Instrument can break at the level of the thread when being exposed to high forces during nail implantation or removal which can lead to complications during nail insertion and removal.

Distribution

Worldwide distribution - US Nationwide distribution and countries of Switzerland, France, Spain, United Kingdom, Italy, Sweden, Netherlands, Australia.

Lot / Code Info

All lots UDI: (01)07613327350210

Root Cause

Under Investigation by firm

Action Taken

Stryker issued Urgent Medical Device Recall Letter (PFA# 2219868) to US consignees on 5-Dec-2019 , letter states reason for recall, health risk and action to take: 1.Please inform users of this Urgent Medical Device Recall and forward this notice to all those individuals who need to be aware within your organization. 2.Immediately check all stock areas and/or operating room storage to determine if any devices from the affected product list are at your facility. A response is required, even though you may not have any physical inventory on site anymore. 3.Quarantine and discontinue use of the recalled devices. 4. Hospitals/Branches/Agencies: Complete and sign the enclosed Urgent Medical Device Recall Business Reply. An OUS customer letter was also issued at the country level.

More recalls by Stryker Gmbh Bohnackerweg 1 Selzach Switzerland

View all recalls by this company →