RecallDepth

Rusch Greenlite Single-Use Metal Laryngoscope Blade for Fiber Optic System

Company
Teleflex Medical Inc
Recall Initiated
January 9, 2020
Posted
February 7, 2020
Terminated
September 11, 2023
Recall Number
Z-1116-2020
Quantity
338,928 (199,018 US and 139,910 OUS) for all products
Firm Location
Morrisville, NC

Reason for Recall

The products are being recalled due to reports of breakage broke at the point of a welded joint either prior to use during setup or during use inside the patient's mouth.

Distribution

Distributed nationwide in the US and to Puerto Rico; OUS to more than 20 countries; To the following government VA medical Centers: VAMC - KANSAS CITY MO 4801 E Linwood Blvd Kansas City MO 64128-2218

Lot / Code Info

Product Code 004551003; Lot Numbers 1811331, 1902331, 1904331

Root Cause

Process control

Action Taken

Letters sent to customers/users to immediately discontinue use and quarantine any products with the affected product code and lot numbers and return them to the manufacturer.