BD FACS Universal Loader (Optional) to the BD FACSLyric Clinical System, Model 662879 (Universal Loader), Rx Only. The firm name on the label is Becton, Dickinson and Company, BD Biosciences, San Jose, CA. The Loader offers various settings to resuspend and mix samples. It can draw from 12 x 75-mm tubes in 30- and 40-tube racks. A barcode reader verifies the ID on tube racks and individual tubes in 30- tube racks. A built-in imaging system provides safety checks, such as verifying the correct rack type and tube layout, and ensures the tubes were loaded correctly.

Company
Becton, Dickinson And Company, Bd Biosciences
Recall Initiated
May 16, 2018
Posted
July 13, 2018
Terminated
August 13, 2018
Recall Number
Z-2598-2018
Quantity
8 devices
Firm Location
San Jose, CA

Reason for Recall

The solenoid plunger tip length is too short resulting in insufficient engagement to the Universal Loader enclosure to lock during use.

Distribution

US Distribution was made to FL, KS, KY, MA, NJ, and SD.

Lot / Code Info

Serial #2662879000001, UDI (01)00382906628794(11)170915(21)Z662879000001;   Serial #2662879000002, UDI (01)00382906628794(11)170915(21)Z662879000002;   Serial #2662879000003, UDI (01)00382906628794(11)170915(21)Z662879000003;   Serial #2662879000004, UDI (01)00382906628794(11)170915(21)Z662879000004;   Serial #2662879000009, UDI (01)00382906628794(11)170915(21)Z662879000009;   Serial #2662879000010; UDI (01)00382906628794(11 )170915(21)Z662879000010;   Serial #2662879000011; UDI (01)00382906628794(11)170915(21)Z662879000011; and   Serial #2662879000012, UDI (01)00382906628794(11)170915(21)Z662879000012.

Root Cause

Nonconforming Material/Component

Action Taken

The firm, BD Life Sciences, produced the "URGENT PRODUCT RECALL-CORRECTION" letter dated 4/6/2018 to its customers during an on-site visit. by the field service engineer made beginning 5/16/2018 to correct the issue. The visits were completed 6/15/2018. If you require further assistance, contact BD Customer/Technical Support BD Customer Support at 877.232.8995 (prompt 2) for instruments in the United States. For customers outside the US, contact your local BD representative or distributor.