Discovery IGS 730 Interventional Fluoroscopic X-ray Systems

This recall has been terminated (originally issued October 5, 2018).

Company
Ge Medical Systems, Scs
Recall Initiated
March 29, 2018
Posted
October 5, 2018
Terminated
June 15, 2021
Recall Number
Z-2386-2018
Quantity
102 devices total
Official Source
View on FDA website ↗

Reason for Recall

The Discovery IGS and Innova IGS systems may experience X-ray abort errors which could result in the loss of imaging mode before or during fluoroscopic use.

Distribution

US distribution: AR, CA, CA, FL, GA, IA, IL, NC, NH, NJ, NM, NY, PA, SC, TN, A,WI, WV. International Distribution: Argentina, Brazil, China, Colombia, Egypt, France, India, Indonesia, Iran, Italy, Japan, Kuwait, Lithuania, Myanmar, Palestine, Poland, Russia, Saudi Arabia, Singapore, Spain, Tunisia, Turkey, United Kingdom, Vietnam.

Lot / Code Info

Model Number 5730890, Serial Numbers (System ID): 00000D3-17-030 (904BHHCHY), 00000D3-17-032 (083172440628), 661327IGS730 (00000D3-18-001), GON4598369 (00000D3-18-004), 082416140028 (00000D3-17-025), M4234253 (00000D3-17-028), YV3601 (00000D3-18-005), 6073XR0058 (00000D3-17-031).

Root Cause

Process control

Action Taken

GE Healthcare notified customers on approximately 03/29/2018, via an URGENT MEDICAL DEVICE CORRECTION letter. Safety instructions included that before each use, to ensure the IGS system is fully functional as indicated in the GE product labeling and follow each facility's established procedures before and during each use to manage the patient if a repeated X-ray abort error occurs during the procedure or in the case a complete loss of fluoroscopic viewing capability may occur. If the problem persists, customers are instructed to contact their local GE Healthcare Service Representative. Additionally, customers are instructed that GE Healthcare will correct all affected products at no cost and a GE Healthcare representative will contact each customer to arrange for the correction. Questions or concerns regarding the notification can be directed to GE Healthcare Service at 1-800-437-1171 or the local Service Representative.