RecallDepth

Evans Plate, Sterile, Rx Only, ORTHOFIX S.r.r. Italy on label, Size (Part Number): a) 6MM Spacer (002-H-00006); b) 8MM Spacer (022-H-00008); c) 10MM Spacer (002-H-00010); d) 12MM Spacer (002-H-00012)

This recall has been terminated (originally issued June 7, 2018).

Company
Orthofix, Inc
Recall Initiated
April 6, 2018
Posted
June 7, 2018
Terminated
June 30, 2021
Recall Number
Z-2127-2018
Quantity
52
Firm Location
Lewisville, TX
Official Source
View on FDA website ↗

Reason for Recall

Firm is conducting a voluntary recall of the Rival Reduce Fracture Plating System and Rival View Plating System. As the result of a complaint investigation, the locking screws were confirmed to not properly engage the plates during surgery.

Distribution

U.S Consignees: NJ, FL, PA, SC, LA, VA, MN, TX, GA, DC, NV, IN, CA, MI, IL, MD, NC, OH, OK,

Lot / Code Info

Lot: B1212604, Date of Manufacture: 2017-11-15, Use By: 2022-10-15, UDI: a) (18056099647724) B1099059, B1123627; b) (18056099647731) B1099057; c) (18056099647748) B1099068; d (18056099647755) B1099069

Root Cause

Device Design

Action Taken

On 04/06/2018, Orthofix Inc. (U.S. Agent), sent written notification of the recall to distributors and hospitals via e-mail and/or FedEx overnight courier.

More recalls by Orthofix, Inc

View all recalls by this company →