H-Plate, Sterile, Rx Only, ORTHOFIX S.r.r. Italy on label, Size (Part Number): a) 12MM Spacer (003-H-12001); b) 15MM Spacer (003-H-15001); 20MM Spacer (003-H-20001); 25MM Spacer (003-H-25001); 30 MM (003-H-30001)

This recall has been terminated (originally issued June 7, 2018).

Company
Orthofix, Inc
Recall Initiated
April 6, 2018
Posted
June 7, 2018
Terminated
June 30, 2021
Recall Number
Z-2128-2018
Quantity
75
Firm Location
Lewisville, TX
Official Source
View on FDA website ↗

Reason for Recall

Firm is conducting a voluntary recall of the Rival Reduce Fracture Plating System and Rival View Plating System. As the result of a complaint investigation, the locking screws were confirmed to not properly engage the plates during surgery.

Distribution

U.S Consignees: NJ, FL, PA, SC, LA, VA, MN, TX, GA, DC, NV, IN, CA, MI, IL, MD, NC, OH, OK,

Lot / Code Info

Lot: B1212400, Date of Manufacture: 2017-11-13, Use By: 2022-10-13, (UDI)/Lot:  a) (18056099647908) B1096740; b) (18056099647915) B1096733; c) (18056099647939) B1096734; d) (18056099647953) B1096735; e) (18056099647960) B1100226

Root Cause

Device Design

Action Taken

On 04/06/2018, Orthofix Inc. (U.S. Agent), sent written notification of the recall to distributors and hospitals via e-mail and/or FedEx overnight courier.