Flexima APDL Drainage Catheter System and Kit, Material Numbers: a) REF M001271960, b) REF M001271970; all-purpose drainage catheter with locking pigtail

Company
Boston Scientific Corporation
Recall Initiated
July 18, 2024
Posted
September 13, 2024
Recall Number
Z-3147-2024
Quantity
89 units
Firm Location
Maple Grove, MN

Reason for Recall

There is the potential for holes in the clear sterile barrier of the device pouch which may impact device sterility.

Distribution

US, Colombia, Philippines, Russia

Lot / Code Info

a) GTIN/UDI 0871472932347, Lot # 33540319, Exp. 03/01/2027; b) GTIN/UDI 0871472932348, Lot # 33471466, Exp. 02/21/2027

Root Cause

Packaging process control

Action Taken

Boston Scientific issued am Urgent Medical Device Removal notice to it consignees on 07/18/2024 via overnight mail. The notice explained the issue, potential hazard, and requested the return of the devices. Consignees were directed to post the notice in a visible location near the affected products to ensure visibility and to share the notice with all those to whom the device was transferred. Distributors were direct to notify their customers.