REF 27219-2 X1, Medtronic, Delta Valve, Neonatal, Performance Level 2.0, CE 0344, Sterile EO - Product Usage: Neonatal Delta Snap Shunt Assemblies are components of CSF-Flow Control Shunt Systems designed to provide controlled CSF flow from the ventricles of the brain into the peritoneal cavity.

Company
Medtronic Neurosurgery
Recall Initiated
December 16, 2020
Recall Number
Z-0872-2021
Firm Location
Irvine, CA

Reason for Recall

Neonatal Valves were manufactured with a ventricular catheter connection site shorter than the design specification. The defect can lead to difficulty in securing the valve to the proximal catheter which could result in disconnection of the ventricular catheter from the valve and shunt malfunction

Distribution

Worldwide distribution - US Nationwide distribution including in the states of AL, AZ, CA, CO, FL, GA, IL, KS, KY, LA, MA, MI, MO, NV, NM, NY, ND, OH, OK, TN, VA and the countries of Albania, Australia, Brazil, Cyprus, Ireland, Mexico, Qatar, Romania, Russian Federation, Serbia, South Africa, Spain, United Kingdom.

Lot / Code Info

Lot # 0219281777

Root Cause

Device Design

Action Taken

On 12/16/2020, Medtronic sent a "URGENT: MEDICAL DEVICE RECALL" notification to affected customers via FedEx. In addition, to notifying consignees about the recall, the firm ask consignees to take the following actions: 1. Identify, segregate, and quarantine affected valves within your inventory. 2. Contact Medtronic to return affected valves and receive replacement(s). See instructions below. 3. Please complete and return the customer confirmation form, even if you do not have any affected valves. 4. Return the form via email to RS.NavFCA@Medtronic.com or via fax to 651-367-7075 within 30 days of receipt.