MEDICREA LIGAPASS TENSION PULLEY HANDLE XS- intended to provide thoracic and/or lumbar posterior stabilization for degenerative pathologies Ref : A08200200

This recall has been terminated (originally issued October 8, 2020).

Company
Medicrea International Vancia Vancia Rillieux La Pape France
Recall Initiated
October 8, 2020
Terminated
May 9, 2024
Recall Number
Z-0463-2021
Quantity
17 units US; 8 units OUS Expanded: 1 US and 3 OUS
Official Source
View on FDA website ↗

Reason for Recall

Tightening the LIGAPASS Band with this instrument, the tensioner's pliers can crush the band causing the band to break during surgery and lead to a risk of infection and re-intervention ,

Distribution

CA Foreign:France, Belgium

Lot / Code Info

Lot Numbers: US: 17E0550-1, 17E1089R/2, 18E0278, OUS: 19B0962, 19B0595, 19B0601  Expanded Recall: 11/3/20 17E0550/2, 19B0602, 19B0963 and 19B0964

Root Cause

Device Design

Action Taken

Medicrea International issued Reportable voluntary Recall notification dated 10/5/20 to via e-mail to distributors and subsidiaries stating reason for recall, health risk and action to take: Instructions sent to US subsidiary : 1. Immediately check your internal inventory and quarantine all subject devices pending return to MEDICREA 2. Circulate this Field Safety Notice internally to all interested/affected parties. 3. Inform MEDICREA if any of the subject devices have been distributed to other organisations. a) Please provide contact details so that MEDICREA can inform the recipients appropriately. b) If you are a Distributor, note that you are responsible for notifying your affected customers. 4. Please inform MEDICREA of any adverse events concerning the use of the subject devices Please comply with any local laws or regulations concerning the notification of adverse events to your National Competent Authority. 5. Complete the attached customer response form. It may be that you no longer have any physical inventory on site. Completing this form will allow us to update our records and will also negate the need for us to send any further unnecessary communications on this matter. Therefore please complete even if you no longer have any of the subject devices in your physical inventory. Contact: Karine TROGNEUX MEDICREA INTERNATIONAL Regulatory Affairs Manager, Direct : +33(0)4.69.85.95.39 Standard : +33 (0)4.72.01.87.87. Fax - N +33 (0)04 72 01 87 88 Mail : ktrogneux@medicrea.com