RecallDepth

LINK SymphoKnee Patella Sizing Template, Item Number 881-509/00

This recall is currently active, issued July 25, 2024. It was issued by Link Bio Corp.

Company
Link Bio Corp
Recall Initiated
June 27, 2024
Posted
July 25, 2024
Recall Number
Z-2445-2024
Quantity
8 units
Firm Location
Dover, NJ
Official Source
View on FDA website ↗

Reason for Recall

The LinkSymphoKnee Patella Sizing Template was incorrectly labeled. The markings on two arms of the instrument were mixed up because the triangular Patella Sizing Template was inserted upside down into the laser marking process.

Distribution

US Nationwide distribution in the states of FL, GA, KS, OH, WI.

Lot / Code Info

Item Number: 881-509/00; UDI-DI: 04026575310203; Lot number: C225066.

Root Cause

Process change control

Action Taken

On June 27, 2024 URGENT: MEDICAL DEVICE RECALL letters were sent via email to distributors. Actions to be taken by the Customer/User: Please check your inventory for this item and lot combination. If the markings are incorrect, as indicated in Section 5 below, please return the affected product to LinkBio. Please respond using the attached Distributor Reply form to indicate your understanding of this Device Recall and of the need to return any affected product. Pre-paid shipping labels will be provided once the Distributor Reply Form is returned, should any affected product be identified. Please return any affected product as soon as possible to the following address: LinkBio Corporation 69 King St Dover, NJ 07801 Please contact LinkBio at 973-625-1333 or complaints@linkbio.com for any questions. Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail or by fax.