RecallDepth

QUIKFLAP, 1X14MM BURR HOLE COVER, 1.5X4mm SELF DRILLING SCREWS, Part Number 12-01538S, for use on the cranium

Company
Stryker Leibinger Gmbh & Co. Kg Botzinger Str. 41 Freiburg Im Breisgau Germany
Recall Initiated
June 22, 2018
Terminated
July 8, 2019
Recall Number
Z-2909-2018
Quantity
22,831 total

Reason for Recall

After a three year term, there is a potential for the external packaging (sterile barrier) to become compromised.

Distribution

US distribution in the states: AZ, FL, NH, and OH, Australia, Canada, France, New Zealand, and UK.

Lot / Code Info

1000167017, 1000167829, 1000171360, 1000172321, 1000173866, 1000175689, 1000178993, 1000187364, 1000198471, 1000198472, 1000202431, 1000205662, 1000215879, 1000217675, 1000226524, 1000233194, 1000238414, 1000259405

Root Cause

Package design/selection

Action Taken

URGENT MEDICAL DEVICE RECALL letters dated 6/22/18 were sent to customers with the following requested actions to be taken: 1. Please inform users of this Medical Device Removal and pass this notice to all appropriate individuals within your organization. 2. Return all affected products available at your location to: Stryker CMF Attn: Recall Coordinator 6300 Sprinkle Rd Portage, MI 49002 REF: PFA #1739180 Or, Contact Stryker Customer Service and refer to PFA #1739180 with any questions about returning the product to us. 3. Complete and sign the enclosed Business Reply Form and fax a copy to: (877) 648-7114 or email a copy to CMF-custserv@stryker.com. 4. Keep a copy of the completed and executed Customer response Form for your records.

More recalls by Stryker Leibinger Gmbh & Co. Kg Botzinger Str. 41 Freiburg Im Breisgau Germany

View all recalls by this company →